    Thermal Mass Flow Meter Calibration Guide

    Time: 2025-09-10 10:09:21 Click:0

    Thermal Mass Flow Meter Calibration Guide

    I. Calibration Process

    1. Zero Calibration Steps

    Power-on operation?: Connect the power supply and allow the meter to operate for 1 hour to several hours with the medium until the device reaches a stable state

    Valve operation?: First close the downstream valve, then close the upstream valve, ensuring the sensor is in a full-pipe state with complete fluid flow stoppage

    Zero adjustment?: Use a hand controller to zero the flow meter, observe zero fluctuations and record

    Stability verification?: Repeat zero adjustment several times until the actual zero fluctuates near zero within a range smaller than the zero stability specification

    2. Range Calibration Methods

    Standard comparison method?: Use a standard flow meter (such as sonic nozzle or bell prover) installed in series with the meter to be calibrated, compare measurements and adjust parameters

    Parameter adjustment?: Adjust parameters such as meter coefficient and segmented velocity correction based on error records

    Accuracy verification?: Verify measurement accuracy using mass or volumetric methods after calibration

    II. Calibration Equipment and Tools

    Equipment Type Working Principle Application Scenario

    Standard flow meter? High-precision measurement (e.g., Coriolis mass flow meter, turbine flow meter) Lab or field calibration, one accuracy grade higher than device under test

    Flow calibrator? Generates known flow (e.g., Defender 500 series) Field calibration, provides stable gas flow output

    Wind tunnel? Produces controlled airflow Lab calibration for large flow/high velocity

    Sonic nozzle? Calculates flow using sonic characteristics High-velocity/large flow calibration (e.g., chemical, petroleum fields)

    Hand controller? Parameter setting and adjustment Field zeroing and parameter setting

    III. Calibration Precautions


    Installation specifications?


    Ensure sufficient straight pipe length: minimum 10 pipe diameters (10D) upstream, 5D downstream

    Flow direction must match meter arrow marking, avoid reverse measurement

    Prohibit low-position installation to prevent water accumulation and debris collection


    Environmental conditions?


    Maintain ambient temperature between 5°C-45°C

    Humidity ≤85%, avoid extreme environments

    Keep away from strong vibration sources and electromagnetic interference


    Medium requirements?


    Prohibit acetylene gas and water/steam-containing media

    Calibration medium must match actual working conditions (gas density, temperature)

    Flow velocity range should be controlled between 0.5-60 m/s


    Safety operations?


    No welding in explosive environments

    Outdoor installations require sun/rain protection covers

    Strictly follow Lockout-Tagout (LOTO) safety procedures

    IV. Common Errors and Solutions

    Error Type Cause Analysis Solution

    Improper installation? Insufficient straight pipe or obstructions Reserve 20D upstream, 10D downstream

    Medium contamination? Water condensation or coating on sensor surface Install upstream filter, regularly clean sensor

    Power issues? Unstable power causes measurement drift Check power indicators, ensure voltage stability